GSG-50-1200
GSG-50-1200 vacuum tube furnace
Vacuum & atmosphere ≤ 1200 °C VacuumInert / reducing Sintering / debindingAnnealingTesting
Part of our tube furnace range — selection notes and all systems →
KEY DATADELIVERED
MAX TEMP
1200℃
ATMOSPHERE
Vacuum / Inert / reducing
CATEGORY
Vacuum & atmosphere
SOURCE
Proposal text
From the original technical proposalAHGSTEC
Specification
Specification
Taken from the original technical proposal. Ranges indicate configurable options; the signed technical agreement governs.
Primary A
- Temperature range
- 100–1200 °C
- Temperature fluctuation
- ±1 °C
- Ultimate vacuum
- -0.1 MPa
- Display resolution
- 1 °C
- Chamber dimensions
- 100 × 600 mm
- External dimensions
- 910 × 410 × 800 mm
- Heated zone length
- 500 mm
- Heating rate
- ≤30 °C/min (freely adjustable)
Secondary B
- Total power
- 4 kW
- Power supply
- 220 V, 50 Hz
- Thermocouple
- Type S thermocouple
- Control system
- 32-segment programmable controller, PID, control accuracy 1 °C
- Tube material
- High-purity quartz / corundum / alloy tube
- Vacuum pump
- Two-stage rotary vane vacuum pump
- Heating elements
- 0Cr27Al7Mo2 alloy wire
Structure
Cross-section
Click any part to see its role. Together these eight determine the final thermal performance.
PART 01
Chamber
One-piece polycrystalline alumina ceramic fiber. High density, low conductivity, no dusting at temperature — the first determinant of uniformity.
- Material
- Polycrystalline alumina fiber
- Continuous rating
- ≤ 650 °C
- Construction
- Seamless, one-piece
Related
Related systems
Debinding and sintering furnace
TEMP1000 °C
ATMOSPHEREAmbient air
PROCESSSintering / debinding
CATEGORYVacuum & atmosphere
GSG-8030-1000
GSG-8030-1000 tube furnace
TEMP1000 °C
ATMOSPHEREAmbient air / Inert / reducing
PROCESSSintering / debinding / Annealing / Testing
CATEGORYVacuum & atmosphere
GSG-40-1100
GSG-40-1100 high-temperature atmosphere tube furnace
TEMP1100 °C
ATMOSPHEREInert / reducing / Vacuum
PROCESSAnnealing / Sintering / debinding / Testing